Art Photonics FlexiSpec® ATR Probes
Type | Window Material | Spectral range | Temperature range | Pressure Max |
---|---|---|---|---|
Reflection | Sapphire, 20° angle | 4000-25000 cm-1 | 0°C / + 150°C | 10 Bar |
Diffuse reflection | Sapphire | 4550-25000 cm-1 | 0°C / + 200°C | 50 Bar |
Side Diffuse reflection | Sapphire | 4550-25000 cm-1 | 0°C / + 200°C | 50 Bar |
Side-Line Diffuse reflection | Sapphire | 4550-25000 cm-1 | 0°C / + 200°C | 50 Bar |
Sapphire window design without straight light Possible design with thinnest probe tips d=0.5 mm Dedicated design providing minimal cross talk due to using of Metal Coated Fibers Flat or angled tip design Round-to-Round or Round-to-Line design Extruder design on request
Type | Fiber | Bundle Arrangement | Spectral range | Temperature range | Pressure Max |
---|---|---|---|---|---|
Polycrystalline | 1+7 | 600-2500 cm-1 | 0°C / + 80°C | 1 Bar | |
Chalcogenide | 1+6 | 1600-7000 cm-1 | 0°C / + 80°C | 1 Bar |
On-line transmission spectroscopy at long distance in wide range of liquids (from transparent to poorly transparent) High throughput in any part of UV - Vis or Vis - NIR spectra Flexible and robust for industrial applications in harsh environment Compatible with all spectrometers ATEX-compatible design on request
Tip material | Spectral range | Temperature range | Pressure Max |
---|---|---|---|
UV-VIS | 260-1200nm | 0°C / + 200°C | 50 Bar |
NIR | 4550-25000cm-1 | 0°C / + 200°C | 50 Bar |
In-line transmission spectroscopy in liquids at long distance High throughput in any part of UV –Vis to NIR spectra Flexible and robust for industrial applications in harsh environment Compatible with all spectrometers and process-interfaces
Tip material | Spectral range | Temperature range | Pressure Max |
---|---|---|---|
UV-Vis | 200-1300nm | -50°C / + 200°C | 200 Bar |
NIR | 4550-25000 cm-1 | -50°C / + 200°C | 10 Bar |
Thinnest probe tips with diameter 0.5 mm Fiber Optic Multi Leg or Y– design Reflection Probe Minimal Cross Talk with metal coated fibers Flat or angled (38o) tip design Round-to-Round, Round-to-Line design
Type | Spectral range | Temperature range | Pressure Max |
---|---|---|---|
UV-Vis | 600-1900 cm-1 | 0°C / + 150°C | 10 Bar |
NIR | 4000-25000 cm-1 | 0°C / + 150°C | 10 Bar |
In-line Raman spectroscopy Single-wavelength excitation (532, 785 nm) Flexible and robust for industrial applications in harsh environment Compatible with all Raman fiber optic spectrometers
Laser Wavelength |
|
---|---|
Laser Spot at the sample |
|
Spectral Range Cut-Off |
|
Filter Efficiency |
|
Laser Transmission | > 80% |
Fiber type |
|
Window Material | Sapphire C-plane AR coated |
Lens Material | Fused Silica AR coated |
Window Sealing Material | Epoxy, Gold, PTFE, Brazing |
Probe Shaft & Body Material | Stainless Steel |
Shaft Size |
|
Fiber Length | 1.2m + 2 legs of 0.5m each |
Total Length (shaft + fiber) | 2.00 ± 0.05m |
Input / Output Connectors | FC/PC or SMA905 |
Protection tubing | Liquid Tight Protection (LTP) |
Operating temperatures | from -20°C to 200°C (up to 300°C on request) |
Pressure (max) | 100 Ba |
Multichannel fiber optic probes NIR and Raman channels can work simultaneously providing hybrid modelling opportunities On-line Raman and NIR spectroscopy Flexible and robust for industrial applications in harsh environment Compatible with all Raman and NIR spectrometers
Probe Shaft | |
---|---|
Probe Shaft: | 170 mm |
Shaft Material: | Stainless Steel 316L |
Shaft OD: | 19 ± 0.1mm |
Total Length: | 3 ± 0.05m |
Temperature range: | 0°C to 150°C |
Pressure (max) | 0 - 10 bar |
Spectral Range: |
|
illumination Bundle: | 19 x NIR400/440, NA=0.22 ± 0.02 |
---|---|
Detection Fiber: | 1 x NIR400/440, NA=0.22 ± 0.02 |
Terminations: | 20 x SMA905 |
Laser Wavelength: | 785 nm |
---|---|
Excitation Fiber: | 1 x NIR105/125, NA=0.22 ± 0.02 |
Detection Fiber: | 1 x NIR200/220, NA=0.22 ± 0.02 |